Short Description

Maintain piping flexibility and velocity accuracy for long time, drastically reduce damage to the liquid ingredients. Unique ripple compensation technology, infusion safer, patients more security.

Description

Details

Technical Features

1.Half Extrusion: Maintain piping flexibility and velocity accuracy for long time, drastically reduce damage to the liquid ingredients.

2.Drip Balance: Unique ripple compensation technology, infusion safer, patients more security.

3.More Infusion Modes: Four infusion modes, operation more convenient and flexible.

4.Original Technology: Original infusion pressure testing technology, more accurate, more safe.

5.Security Design: Unique security design, multiple monitoring technology, ensuring transfusion safety, even CPU is damaged still can stop infusion and alarm.

6.Operation Record: Operating history record, access to convenient, and will not be deleted.

Main Technical Specification

1.Pump Mode: Half extrusion finger peristaltic pump

2.Applicable Infusion: Outer diameter 3.4 mm ~ 4.5 mm, high quality PVC, TPE infusion set.

3.Four Infusion Modes: Flow –rate mode, time-amount mode, drop-rate mode, body-weight mode

4.Flow-rate Mode: Flow-rate 0.1-1200ml/h, 0.1 ml/h step

5.Time-amount Mode: Time 1s~99h 59m 59s. 1s step

6.Drop-rate Mode: Drop-rate 1~333d/min. 1 d/min step

7.Body Weight Mode:
Does: 0.1~999.9 mg, 0.1 mg step
mg/kg/h: 0.01~999.99 mg/kg/h, 0.01 mg/kg/h step
Body-weight: 1~300 kg, 1 kg step

8.KVO Function: KVO rate 1.0~5.0 ml/h, 0.1 ml/h step

9.Bolus Function:
Bolus rate: 50.0~800.0 ml/h, 0.1 ml/h step
Bolus volume: 0.1~99.9 ml/h, 0.1 ml/h step

10.Infusion Precision: Better than ±5%(using a calibrated or a special infusion set)

11.Alarm:
Air in tube, 3 grades of sensitivity.
Bubble sensor fault
Upper-end occlusion, -20~-60 kPa, 5 grades of sensitivity
Upper-end base-pressure fault
Lower-end occlusion, 20-140 kPa, 5 grades of sensitivity
Lower-end base-pressure fault
Pump rotation fault
Inverse-infusion
Abnormal door opening
Low battery power, battery failure
Overtime pause
Completed infusion
Ac power fault
MCU fault, and stop infusion

12.Data Function:
Memory of calibration: Recodes of calibration data of tube automatically for next operation.
Parameter memory function: Automatic recording the last infusion parameters, to facilitate follow-up using.
History function: Can record recently 1000 infusion events(what/when/why, flow-rate,etc.)
Power Supply: Ac network power supply: 100 ~ 240V, 50/60 Hz, built-in rechargeable nimh battery.

13.Power Consumption: ≤20VA
14.Size: 130 mm * 206 mm * 118 mm(W*H*T)
15.Weight: 1.3 kg